The Simple Logger II Model L702 is a two-channel Temperature/Humidity recording device powered by alkaline batteries. It is capable of measuring temperatures from 14 to 122°F (-10 to 50°C) and 5 to 85% RH with resolutions of 0.1 degrees and 0.1% RH.
Model L702 stores measurements at user-programmable rates of once every 5 seconds to 1 every day. Up to 240,000 measurements can be stored for a period of 45 days or more.
The L702 incorporates a unique capacitive sensor element for measuring relative humidity while the temperature is measured by a band-gap sensor. CMOSens technology guarantees excellent reliability and long-term stability with high accuracy.
Analog information on the input is sampled and converted to a digital signal with up to 14-bit resolution. An optically isolated Universal Serial Bus (USB) port provides for the transfer of data from the instrument’s internal memory to the computer for analysis.
One of the main advantages of the logger is its ability to perform a variety of recording tasks with an easy and intuitive setup from a computer using DataViewsoftware. DataView is included with the L702 and provides a real-time display of measurements, data analysis, and report generation.
Applications:
- Clean Rooms
- Blood Banks
- Humidors
- Wine Cellars
- Greenhouses
- Paper Mills
DataView Graphing & Analysis Software
The DataView software provides a convenient way to configure and control logging from your computer. Through the use of clear and easy-to-use tabbed dialog boxes, all Simple Logger II functions can be configured and tests can be initiated. Results can be displayed in real-time and stored in your PC or the logger. Standard or custom reports may be printed along with the operator’s comments and analysis.
Software Features
- Display and analyze real-time data on your PC
- Configure all Data Logger functions and parameters from your PC including sample rate, recording length, channel configuration, and more
- Select storage mode for optimum recording
- Zoom in and out and pan through sections of the graph to analyze the data
- Download, display, and analyze recorded data
- Display waveforms, trend graphs, harmonics (AC models), and text summaries, all in real-time
- Create custom views and reports
- Print reports using standard or custom templates you design
- Free software upgrades
3 Recording Modes
Start/Stop Mode
In this mode, the operator selects a storage rate from predefined values from 8 per second to 1 every day, then selects a start and stop time. Data is recorded at this rate until the memory is filled or the end recording time and date is reached. The logger then stops recording and goes into standby mode retaining the recorded date.
First In First Out Mode (FIFO)
The operator makes the data storage and recording length selections as described above. If the memory fills before the end date and time occurs, the logger will discard the oldest stored data point and add a new one. This process will continue until the end recording date and time is reached.
Extended Time Recording Mode (XRM)
This unique recording mechanism provides for continuous recording over a longer period of time without the need for the operator's selection of adjustments. The operator selects a starting storage range from predefined values from 8 per second to 1 every day and a recording length. The logger will store data at the rate selected until the memory is filled. The logger will then discard every other stored sample beyond the first one, freeing up half the memory for continuous recording. New samples will be stored at half the previous storage rate, matching the interval with the remaining stored data. This process will repeat each time the memory fills until the operator stops the recording manually, the end recording date and time is reached or the battery runs down.