AEMC 2114.95 Specifications |
Measurement Range |
Range |
0.00 to 19.99W / 20.00 to 199.9W / 200.0 to 1999W |
Measurement Range |
20W / 200W / 2000W |
Resolution |
10mW / 100mW / 1W |
Test Current |
10mA / 1mA / 0.1mA |
Accuracy in % of Reading |
2% Reading ±1 ct / 3% Reading ±3 cts |
Accuracy Reference Conditions |
23°C ±3K, 50% RH ±10%, battery at 8V ±0.2V, externa; magnetic field <40A/m, external field <1 V/m, conductor centered, loop resistance noninductive. Accuracy % of reading |
Resistance Measurement Frequency |
128Hz square wave |
Interference |
Model 4610Kit rejects high levels of interference voltage (DC, 50 to 60Hz, harmonics); DC voltage in series with X:20V AC voltage in series with Y:13V peak AC voltage in series with Z:32V peak |
Response Time |
Approximately 6 seconds for a stabilized measurement |
Voltage Withstand |
250VACc with spikes of 3000VAC or 1000 |
Power Source |
Eight 1.5V "AA" batteries; Alkaline recommended;"LO BAT" indication on LCD |
Battery Life |
1800 15-second measurements |
Fuse Protection |
High breaking capacity 0.1A, 250V, 0.25" x 1.25" |
Mechanical Specifications |
Display |
7 segment LCD, .71" (18mm) high (3-1/2 digit); 2000 counts; LCD also indicates overrange, test lead shorts and lead reversals |
Connection |
Color-coded terminals accept lugs with min. gap of 6mm or standard 4mm banana jacks |
LED Indication |
3 LEDs indicate high input noise, high auxiliary rod resistance, open leads, blown fuse |
Operating Temperature |
14° to 131°F (-10° to 55°C),0 to 90%RH |
Storage Temperature |
-40° to 158° F (-40 to 70° C), 0 to 90% RH with batteries removed |
Dimensions |
8.7 x 5.4 x 5.9" (220 x 136 x 150mm) |
Weight |
10 lbs. |
Case |
Heavy-duty, ABS |
Colors |
Case: safety yellow; Front panel: gray |
Mechanical Shock |
IEC 68-2-27 |
Vibration Test |
IEC 68-2-6 |
Drop Test |
IEC 68-2-32 |
Dielectric Test |
3kV, 50/60Hz, 1 min. between four interconnected measuring terminals and any external metal ground |
Environmental |
O-ring sealed against dust and water to IP50 (Protection Index) |
Electrostatic |
IEC801-2 |
Electromagnetic |
IEC 801-3 |
Electric Shcok |
IEC 801-5 |
Safety Reference Specifications |
Safety Ratings |
IEC 1010-1 Cat III Pollution Degree 2, 42V |
Agency Approvals |
Emission: EN 50081-1; Immunity: EN 50082-1 |